MEAS/SOURCE switches direction, V / Ω / TC / RTD select the function, and the arrow keys set the value. A technician can use it correctly without training, which is the whole point of a field calibrator.| Model | Temperature | Pressure | Frequency | HART | Documenting |
|---|---|---|---|---|---|
| Fluke 724 | Measure + source, 12 TC + 7 RTD | No | No | No | No |
| Fluke 725 | Yes | Yes, via 700P series modules | Yes | No | No |
| Fluke 726 | Yes | Yes, via 700P series modules | Yes | Yes | Yes, with memory and PC interface |
Source a simulated RTD or thermocouple into the transmitter's sensor input, power the loop from the 724's 24 V supply, and read the 4–20 mA output on the upper display. One connection, one operator, both numbers visible.
Measure the actual RTD or thermocouple and compare it with what the control system reports. Half of all "transmitter drift" calls are a sensor that has drifted, and this separates the two in a single visit.
Five-point checks that used to be a manual exercise now run as a step function while the technician records the loop current.
The 724 is specified for high-speed pulsed RTD circuits, which is exactly where cheap simulators fail on older transmitters.
The dual display and the dedicated LOOP key make a 4–20 mA loop check a two-minute job rather than a two-tool job.
0.02 % + 2 counts on volts and mA, plus RTD accuracy around 0.2 °C, is enough for in-house verification of most plant instrumentation — though not for accredited calibration of a reference standard.
Food and beverage (pasteurisation and CIP loops), pharmaceutical (validated temperature loops), power generation (RTD and TC instrumentation), oil and gas (transmitter maintenance), HVAC and building services (chilled water and AHU sensors), metals and heat treatment (furnace thermocouples).
Buy the 724 if your work is temperature: RTDs, thermocouples, and the 4–20 mA transmitters attached to them. Buy the 725 if you also need pressure, frequency or a wider electrical range, because it accepts the 700P series pressure modules. Buy the 726 if you need to document the results — it stores readings and talks to a PC, which is what an audit trail requires. The 724 has no pressure capability and cannot be upgraded to one.
Only for Ni120, Pt200 and Pt1000. The full picture: RTDs run 0.2 to 0.33 °C, and thermocouples run 0.7 to 1.8 °C depending on type. Type J is 0.7 °C, type K is 0.8 °C, type R is 1.8 °C. Marketing material that says "0.2 °C accuracy" is quoting the best RTD in the table as though it applied to everything. If you publish it that way, the first customer who checks a type K against a dry block will call it a fault.
Yes. The 724 sources the sensor signal, supplies 24 V loop power, and measures the loop current simultaneously on the upper display, with 25 % and 100 % step functions and auto ramp for the sequence. It is a complete single-tool loop calibration for a temperature transmitter.
Yes, and the accuracy differs on the lowest range: 0.1 Ω on a 4-wire connection, 0.15 Ω on 2- and 3-wire. Measure with four wires where you can — that is what the published RTD temperature accuracies are based on.
No. The 724 stores test setups for recall; it does not store readings and it has no PC interface. If you need recorded results, an export file, or an audit trail, that is the 726 documenting calibrator.
No. There is no pressure function and no module bay. Pressure calibration needs the 725 or 726 plus a 700P series pressure module. This is the single most common mismatch between what a buyer expects from a "process calibrator" and what the 724 delivers, so it is worth stating on the page rather than in the returns email.