Fluke 923 Hot-Wire Anemometer - Detachable Probe Air Velocity Meter for Cleanroom & Duct Testing

Fluke 923 Hot-Wire Anemometer - Detachable Probe Air Velocity Meter for Cleanroom & Duct Testing

  • Highlight

    Hot-Wire Anemometer Detachable Probe

    ,

    Air Velocity Meter Cleanroom Testing

    ,

    Duct Testing Anemometer

  • Air Velocity Range
    0.20 To 20.00 M/s (40 To 3940 Fpm), Resolution 0.01 M/s
  • Sensor Type
    Hot-wire Thermal, Slim Telescopic Probe
  • Probe Connection
    Detachable - Infrared When Attached, RF Wireless When Separated
  • Air Flow Range
    0 To 99,999 M3/hr, Cfm Or L/s, Computed From Duct Dimensions
  • Temperature Range
    -20 To 60 C, +/-0.5 C From 0 To 45 C
  • Stored Samples And Alarm
    198 Samples (99 Velocity + 99 Flow); Velocity Alarm With Vibration And Beep
  • Shipping Term
    International Express(DHL/FEDEX/TNT/UPS/ARAMEX)/Sea Freight / Air Freight / International Special Line
  • Freight Charge
    For Industrial Equipment (E.G., PLC Modules, Inverters..), Shipping Cost Varies By Package Weight/Volume And Delivery Location. Contact Our Sales Team For A Customized Shipping Solution And Quot
  • Brand Name
    Fluke
  • Certification
    IEC/EN 61326-1 (EMC)、CE、
  • Model Number
    Fluke 923
  • Minimum Order Quantity
    1
  • Price
    $450–529
  • Packaging Details
    Original
  • Delivery Time
    3-5Workdays
  • Payment Terms
    L/C,D/A,D/P,T/T

Fluke 923 Hot-Wire Anemometer - Detachable Probe Air Velocity Meter for Cleanroom & Duct Testing

Product Highlights
  • Reads down to 0.20 m/s. A hot-wire thermal sensor holds accuracy where vane anemometers stall — cleanroom FFUs, small outlets and duct slits.
  • Detachable, telescopic probe. Use it clipped to the meter or lift it off and work in wireless mode; the display stays where you can read it.
  • Velocity, airflow and temperature from one instrument. Flow is computed from duct type and dimensions on board — circular diameter, or rectangular X and Y.
  • Slim probe for tight access. The sensor head fits grilles, diffusers and narrow openings that a 70 mm vane cannot enter.
  • 198 stored samples — 99 velocity plus 99 flow readings, recallable on screen.
  • Average, min/max and data hold across velocity, temperature and flow.
  • Velocity alarm that you can feel. When the set threshold is crossed, the base vibrates and beeps once per second — useful when the display is out of sight.
  • Backlit LED display for plant rooms and ceiling voids.
  • MIL-T-28800F Class 2 vibration and a 1 m drop rating.
  • Two-year warranty from Fluke.
Specifications
Function Range Resolution Accuracy
Air velocity, m/s 0.20 to 20.00 m/s 0.01 m/s ±5 % + 3 digits of reading, or ±1 % + 1 digit of full scale
Air velocity, fpm 40 to 3940 fpm 1 fpm ±5 % + 6 digits of reading, or ±1 % + 1 digit of full scale
Air flow 0 to 99,999 m³/hr, cfm, l/s 0.001 to 100 Function of velocity and duct size
Air temperature -20 °C to 60 °C (-4 °F to 140 °F) 0.1 °C / 0.1 °F ±0.5 °C from 0 °C to 45 °C
Air temperature — — ±1.0 °C from -20 °C to 0 °C and 45 °C to 60 °C
Air temperature, °F — — ±1.0 °F from 32 °F to 113 °F; ±2.0 °F from -4 °F to 32 °F and 113 °F to 140 °F
Applications
  • Cleanroom FFU and terminal filter testing — the low-velocity range that vane meters cannot resolve.
  • Small air outlets and slot diffusers — slim probe enters where a large vane head will not.
  • Duct slits and narrow openings — spot checks where a full traverse is not practical.
  • Fume hood and biological safety cabinet face velocity — repeatable readings at the 0.3 to 0.6 m/s working band.
  • HVAC commissioning and air balancing — velocity plus computed flow from duct dimensions.
  • Laboratory and pharmacy ventilation checks — room supply and exhaust verification.
  • Electronics and semiconductor fab ventilation — airflow at equipment intakes and mini-environments.
  • Drying, cooling and process air — speed across product in ovens and cooling tunnels.
  • Greenhouse and agricultural ventilation — low-speed air exchange measurement.
  • Building services troubleshooting — tracing a weak outlet back to its branch.
Frequently Asked Questions
Q1. Does the 923 measure barometric pressure?

No — it lets you enter it.

This is the single most misquoted point about this model. Barometric pressure is a required input setting (shown in mmHg, defaulting to sea level) that the meter uses to compensate the velocity calculation. There is no pressure sensor, no pressure port and no differential pressure reading. If a job needs actual pressure measurement, that is a different instrument — the Fluke 922 airflow meter.

Q2. Hot-wire or vane — which one do I need?

Hot-wire for slow air in small spaces; vane for fast air in large openings.

The 923's range starts at 0.20 m/s because a thermal sensor resolves air movement that would not turn a vane at all. That makes it the right tool for FFUs, hood faces and slot diffusers. A vane meter covers higher speeds and larger ducts more cheaply, but it reads nothing useful below roughly 0.4 m/s.

Q3. How does the detachable probe work?

Two links, automatically selected.

With the probe seated on the meter body, the two halves talk over infrared. Lift the probe off and it switches to radio frequency, indicated by a flashing wireless symbol. If the RF link does not establish, reseat the probe and power cycle the meter — the manual's own remedy.

Q4. How accurate is it at the bottom of the range?

±5 % of reading + 3 digits, or ±1 % of full scale + 1 digit.

The percentage-of-full-scale clause is what protects low speeds: it caps the error band instead of letting a percentage-of-reading term balloon as the value approaches zero. Add 0.1 × accuracy for each degree outside 18 to 28 °C.

Q5. What batteries does it take?

Two different types — 3 × AAA in the base, 1 × 9 V in the probe.

The probe is battery-powered precisely because it detaches. Fluke does not publish a runtime figure for this model, so carry spares rather than planning around a number. The meter also has separate low-battery indicators for base and probe, so watch both.

Q6. What is in the box, and what is the warranty?

Meter, detachable telescopic probe, probe sensor cover, protective case, batteries and manual. Two-year Fluke warranty.