| Item | 961A | 961B | 961C |
|---|---|---|---|
| Sensor type | Internal | External probe | External probe |
| Temperature range | −30.0 to 70.0 °C | −85.0 to 150.0 °C | −40.0 to 85.0 °C |
| Temperature accuracy | ±0.4 °C | ±(0.3 + 0.005 × |reading|) °C | ±0.4 °C |
| Temperature resolution | 0.1 °C | 0.1 °C | 0.1 °C |
| Humidity range | 0 to 100.0 %RH | Not available | 0 to 100.0 %RH |
| Humidity accuracy | ±3 %RH (20–80 %RH @ 25 °C); ±5 %RH (<20 %RH, >80 %RH @ 25 °C) | — | ±3 %RH (20–80 %RH @ 25 °C); ±5 %RH (<20 %RH, >80 %RH @ 25 °C) |
| Humidity resolution | 0.1 %RH | — | 0.1 %RH |
| Weight (with battery) | 60 g | 88 g | 78 g |
| Best for | Fridges, cabinets, ambient rooms | Ultra-low freezers, autoclaves, ovens | Cold rooms, chillers, pharma storage |
No — the PDF report is generated by the logger itself. Plug the USB connector into any Windows PC and a formatted PDF report opens, with min/max/average, alarm events and the full trace. FlukeLog desktop software is only needed when you want Excel or TXT export, custom PDF templates, chart views or remote configuration of start time, interval and alarm limits.
Sensor placement and range, not data features. The 961A has an internal sensor for −30 to 70 °C plus humidity. The 961B has an external probe reaching −85 to 150 °C and no humidity channel. The 961C has an external probe for −40 to 85 °C plus humidity. All three share the same 32,000-point memory, 10 s to 24 hr interval, USB PDF output and IP65 housing.
No. The 961A is rated −30 to 70 °C only. Putting an internal-sensor logger into a ULT freezer exceeds its range and voids the reading. For −80 °C work you need the 961B with its external probe, which covers −85 to 150 °C. This is also why the logger body itself is only rated −30 to 70 °C on every model — the body stays outside, the probe goes in.
Yes — the data path is FDA 21 CFR Part 11 compliant and traceable. Reports carry device identity and timestamps, alarms are recorded as single or cumulative events, and readings cannot be edited after the fact. That is the combination auditors ask for under GDP/GMP and HACCP. Note this is a data-integrity compliance statement, not a calibration certificate.
One year, and yes — it is a user-replaceable CR2450 coin cell. The one-year figure assumes a 15-minute logging interval at 25 °C. Logging every 10 seconds in a cold room will shorten it; estimate roughly proportional to interval. Swap the cell and the stored data is retained.
±0.4 °C on 961A and 961C; ±(0.3 + 0.005 × |reading|) °C on 961B. Resolution is 0.1 °C and 0.1 %RH. The published figures are manufacturer specifications; the datasheet does not state an accredited calibration certificate in the box, so if your SOP requires ISO 17025 traceability, budget for a third-party calibration at purchase.