| Model | Range | Resolution | Accuracy | Burst pressure |
|---|---|---|---|---|
| Fluke-700G27 | −12 to 300 psi / −0.83 to 20 bar | 0.01 psi / 0.001 bar | Positive pressure ±0.05 % FS, vacuum ±0.1 % FS | 1200 psi / 80 bar |
| Fluke-700G29 | −14 to 3000 psi / −0.97 to 200 bar | 0.1 psi / 0.01 bar | Positive pressure ±0.05 % FS, vacuum ±0.1 % FS | 10,000 psi / 690 bar |
| Fluke-700G30 | −14 to 5000 psi / −0.97 to 340 bar | 0.1 psi / 0.01 bar | Positive pressure ±0.05 % FS, vacuum ±0.1 % FS | 10,000 psi / 690 bar |
| Fluke-700G31 | −14 to 10,000 psi / −0.97 to 690 bar | 1 psi / 0.01 bar | Positive pressure ±0.05 % FS, vacuum ±0.1 % FS | 20,000 psi / 1035 bar |
| Model | Full scale | ±0.05 % FS in psi | Error as % of reading at 10 % FS | Error as % of reading at 50 % FS |
|---|---|---|---|---|
| 700G27 | 300 psi | ±0.15 psi | 0.5 % | 0.1 % |
| 700G29 | 3000 psi | ±1.5 psi | 0.5 % | 0.1 % |
| 700G30 | 5000 psi | ±2.5 psi | 0.5 % | 0.1 % |
| 700G31 | 10,000 psi | ±5 psi | 0.5 % | 0.1 % |
| Item | Specification |
|---|---|
| Pressure connection | 1/4 in NPT male |
| Media compatibility | 100, 300, 500, 1000 psi ranges: any liquid or gas compatible with 316 stainless steel. Above 1000 psi: any non-flammable, non-toxic, non-explosive, non-oxidising liquid or gas compatible with 316 stainless steel |
| Display | 5½ digits, 16.53 mm digit height, 20-segment bar graph 0–100 % |
| Power | 3 × AA alkaline batteries |
| Battery life | 1,500 hours continuous with backlight off; 2,000 hours at slow sample rate |
| Dimensions (H × W × D) | 12.7 × 11.4 × 3.7 cm (5 × 4.5 × 1.5 in) |
| Housing and weight | Cast ZNAL housing, 0.56 kg (1.22 lb) with holster |
| Data logging | Up to 8,493 pressure measurements to memory — requires 700G/TRACK software and cable, sold separately |
| Warranty | Three years (Fluke published) |
Take the gauge to the instrument, connect to the process tap or a hand pump, and compare indicated against applied. The 0.05 % FS accuracy gives you roughly a 4:1 test uncertainty ratio against a 0.5 % process gauge, which is what most quality procedures ask for.
Use the 700G as the pressure reference in a transmitter calibration loop: apply pressure, read the reference on the 700G, read the mA output on a calibrator. Pair it with the 700PTPK for pneumatic work to 600 psi or the 700HTPK for hydraulic work to 10,000 psi.
The Min/Max capture holds the actuation and reset points, so you can set a switch and read the trip point afterwards without watching the display during the ramp.
Static proof testing of vessels, hoses and fittings, with the 700G29/30/31 covering the higher test pressures. The 3:1 burst margin on the 700G29 and 700G31 gives working headroom over the test pressure.
With the optional 700G/TRACK software and cable, log up to 8,493 readings and review pressure decay over time instead of watching a display for ten minutes.
Set and verify relief valve cracking pressure, accumulator pre-charge and pump deadhead pressure on the 3000–10,000 psi range, where an analogue gauge's parallax and 1.6 % class make the number meaningless.
Buy the lowest range that covers your highest working pressure, plus roughly 20 % headroom. Accuracy here is a percentage of full scale, so the error in psi scales with the range you choose: the 700G31 is ±5 psi anywhere on its dial, which is 1 % of reading if you mostly work at 500 psi, while the 700G27 is ±0.15 psi at any point in its 300 psi span. If you test at both 100 psi and 8000 psi, that is two gauges, not one.
No — it is a precision test gauge. Fluke's reference-class instruments are the 700RG models, specified at ±0.04 % of reading + 0.01 % of range, which is materially better than ±0.05 % of full scale. If your procedure or accreditation scope requires a reference-class standard, ask for the 700RG29/30/31 instead. Selling the 700G as a "standard" is the kind of over-claim that gets caught at audit.
No. It is a stand-alone gauge with its own display, keypad, housing and batteries — you read it directly, with no calibrator required. A pressure module (Fluke 700P or 750P series) has no display and only works when plugged into a compatible calibrator. The Chinese listing description calls this a "pressure module"; it is not one, and the distinction matters because the two are not interchangeable.
No. The gauge can hold 8,493 measurements internally, but you need the 700G/TRACK software and communication cable to configure the logging run and to upload the results — that is a separately purchased accessory (list around 720–739). Without it, the gauge is a very good indicating gauge and nothing more. Say this before the sale, not after.
On the 300, 500 and 1000 psi ranges, any liquid or gas compatible with 316 stainless steel. Above 1000 psi — which covers the 700G29, 700G30 and 700G31 — Fluke restricts the media to non-flammable, non-toxic, non-explosive and non-oxidising liquids or gases compatible with 316 stainless steel. Oxygen service, hydrogen and flammable refrigerants are therefore out on the high ranges. The port is 1/4 in NPT male, with a 1/4 in NPT to 1/4 in ISO adapter included.
No case. The carton contains the gauge, a traceable certificate of calibration, the 1/4 in NPT to 1/4 in ISO adapter, and the manual on CD-ROM in 14 languages.